Abstract
The performance of a rapidly swept external cavity quantum cascade laser (ECQCL) system combined with an open-path Herriott cell was evaluated for time-resolved measurements of chemical species with broad and narrow absorption spectra. A spectral window spanning 1278-1390 cm-1 was acquired at a 200 Hz acquisition rate, corresponding to a tuning rate of 2×104 cm-1/s, with a spectral resolution of 0.2 cm-1. The capability of the ECQCL to measure < 100 ppbv changes in nitrous oxide (N2O) and 1,1,1,2-tetrafluoroethane (F134A) concentrations on millisecond timescales was demonstrated in simulated plume studies with releases near the open-path Herriott cell. Absorbance spectra measured using the ECQCL system exhibited noise-equivalent absorption coefficients of 5×10-9 cm-1Hz-1/2. For a spectrum acquisition time of 5 ms, noise-equivalent concentrations (NEC) for N2O and F134A were measured to be 70 and 16 ppbv respectively, which improved to sub-ppbv levels with averaging to 100 s. Noise equivalent column densities of 0.64 and 0.25 ppmv × m in 1 sec are estimated for N2O and F134A.
Original language | English (US) |
---|---|
Pages (from-to) | 25553-25569 |
Number of pages | 17 |
Journal | Optics Express |
Volume | 23 |
Issue number | 20 |
DOIs | |
State | Published - Oct 5 2015 |
Externally published | Yes |