T1 - Design of a practical model-observer-based image quality assessment method for x-ray computed tomography imaging systems
AU - Tseng, Hsin Wu
AU - Fan, Jiahua
AU - Kupinski, Matthew A.
N1 - Funding Information: This project was supported by GE Healthcare, Waukesha, Wisconsin. Publisher Copyright: © 2016 Society of Photo-Optical Instrumentation Engineers (SPIE).
PY - 2016/7/1
Y1 - 2016/7/1
N2 - The use of a channelization mechanism on model observers not only makes mimicking human visual behavior possible, but also reduces the amount of image data needed to estimate the model observer parameters. The channelized Hotelling observer (CHO) and channelized scanning linear observer (CSLO) have recently been used to assess CT image quality for detection tasks and combined detection/estimation tasks, respectively. Although the use of channels substantially reduces the amount of data required to compute image quality, the number of scans required for CT imaging is still not practical for routine use. It is our desire to further reduce the number of scans required to make CHO or CSLO an image quality tool for routine and frequent system validations and evaluations. This work explores different data-reduction schemes and designs an approach that requires only a few CT scans. Three different kinds of approaches are included in this study: a conventional CHO/CSLO technique with a large sample size, a conventional CHO/CSLO technique with fewer samples, and an approach that we will show requires fewer samples to mimic conventional performance with a large sample size. The mean value and standard deviation of areas under ROC/EROC curve were estimated using the well-validated shuffle approach. The results indicate that an 80% data reduction can be achieved without loss of accuracy. This substantial data reduction is a step toward a practical tool for routine-task-based QA/QC CT system assessment.
AB - The use of a channelization mechanism on model observers not only makes mimicking human visual behavior possible, but also reduces the amount of image data needed to estimate the model observer parameters. The channelized Hotelling observer (CHO) and channelized scanning linear observer (CSLO) have recently been used to assess CT image quality for detection tasks and combined detection/estimation tasks, respectively. Although the use of channels substantially reduces the amount of data required to compute image quality, the number of scans required for CT imaging is still not practical for routine use. It is our desire to further reduce the number of scans required to make CHO or CSLO an image quality tool for routine and frequent system validations and evaluations. This work explores different data-reduction schemes and designs an approach that requires only a few CT scans. Three different kinds of approaches are included in this study: a conventional CHO/CSLO technique with a large sample size, a conventional CHO/CSLO technique with fewer samples, and an approach that we will show requires fewer samples to mimic conventional performance with a large sample size. The mean value and standard deviation of areas under ROC/EROC curve were estimated using the well-validated shuffle approach. The results indicate that an 80% data reduction can be achieved without loss of accuracy. This substantial data reduction is a step toward a practical tool for routine-task-based QA/QC CT system assessment.
KW - CHO
KW - CSLO
KW - data reduction
KW - leave-one-out likelihood
KW - model observer
UR - http://www.scopus.com/inward/record.url?scp=85006226820&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85006226820&partnerID=8YFLogxK
U2 - 10.1117/1.JMI.3.3.035503
DO - 10.1117/1.JMI.3.3.035503
M3 - Article
SN - 2329-4302
VL - 3
JO - Journal of Medical Imaging
JF - Journal of Medical Imaging
IS - 3
M1 - 035503
ER -