Abstract
Confocal fluorescence ratiometric imaging is an optical technique used to measure a variety of important biological parameters. A small amount of chromatic aberration in the microscope system can introduce a variation in the signal ratio dependent on the fluorophore concentration gradient along the optical axis and lead to bias in the measurement. We present a theoretical model of this effect. Experimental results and simulations clearly demonstrate that this error can be significant and should not be ignored.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 95-102 |
| Number of pages | 8 |
| Journal | Applied optics |
| Volume | 50 |
| Issue number | 1 |
| DOIs | |
| State | Published - Jan 1 2011 |
ASJC Scopus subject areas
- Atomic and Molecular Physics, and Optics
- Engineering (miscellaneous)
- Electrical and Electronic Engineering
Fingerprint
Dive into the research topics of 'Errors in confocal fluorescence ratiometric imaging microscopy due to chromatic aberration'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS