Abstract
A high-performance liquid chromatographic (HPLC) assay suitable for the evaluation of 3-methoxy-4-hydroxyphenylglycol (MHPG) in 1-ml aliquots of plasma is presented. Preparatory steps include extraction into ethyl acetate and minicolumn chromatography. Recoveries are monitored with [3H]MHPG. The HPLC procedure utilizes a C18 column, isocratic elution and amperometric detection. The assay was checked against a gas chromatographic-mass spectrometric procedure; the two procedures correlated well with a correlation coefficient of 0.99. Intra-assay reproducibility was 5.4%, inter-assay reproducibility 10.7%. Immediate changes in the orthostatic position did not affect the plasma MHPG concentration. Based on 22 normal controls the daytime plasma MHPG level was 2.98 ± 0.66 ng/ml (mean ± S.D.).
| Original language | English (US) |
|---|---|
| Pages (from-to) | 91-99 |
| Number of pages | 9 |
| Journal | Journal of Chromatography B: Biomedical Sciences and Applications |
| Volume | 377 |
| Issue number | C |
| DOIs | |
| State | Published - 1986 |
| Externally published | Yes |
ASJC Scopus subject areas
- General Chemistry
Fingerprint
Dive into the research topics of 'Evaluation of plasma 3-methoxy-4-hydroxyphenylglycol'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX