Abstract
In this paper, a four-point characterization method is developed for samples that have either capacitive or ohmic contacts. When capacitive contacts are used, capacitive current- and voltage-dividers result in a capacitive scaling factor not present in four-point measurements with only ohmic contacts. From a circuit equivalent of the complete measurement system, one can determine both the measurement frequency band and capacitive scaling factor for various four-point characterization configurations. This technique is first demonstrated with a discrete element four-point test device and then with a capacitively and ohmically contacted Hall bar sample over a wide frequency range (1 Hz-100 kHz) using lock-in measurement techniques. In all the cases, data fit well to a circuit simulation of the entire measurement system, and best results are achieved with large area capacitive contacts and a high input-impedance preamplifier stage. An undesirable asymmetry offset in the measurement signal is described which can arise due to asymmetric voltage contacts.
| Original language | English (US) |
|---|---|
| Article number | 024703 |
| Journal | Review of Scientific Instruments |
| Volume | 83 |
| Issue number | 2 |
| DOIs | |
| State | Published - Feb 2012 |