Skip to main navigation Skip to search Skip to main content

Predicting ablation zones with multislice volumetric 2-D magnetic resonance thermal imaging

  • Zahabiya Campwala
  • , Benjamin Szewczyk
  • , Teresa Maietta
  • , Rachel Trowbridge
  • , Matthew Tarasek
  • , Chitresh Bhushan
  • , Eric Fiveland
  • , Goutam Ghoshal
  • , Tamas Heffter
  • , Katie Gandomi
  • , Paulo Alberto Carvalho
  • , Christopher Nycz
  • , Erin Jeannotte
  • , Michael Staudt
  • , Julia Nalwalk
  • , Abigail Hellman
  • , Zhanyue Zhao
  • , E. Clif Burdette
  • , Gregory Fischer
  • , Desmond Yeo
  • Julie G. Pilitsis

Research output: Contribution to journal › Article › peer-review

Abstract

Background: High-intensity focused ultrasound (HIFU) serves as a noninvasive stereotactic system for the ablation of brain metastases; however, treatments are limited to simple geometries and energy delivery is limited by the high acoustic attenuation of the calvarium. Minimally-invasive magnetic resonance-guided robotically-assisted (MRgRA) needle-based therapeutic ultrasound (NBTU) using multislice volumetric 2-D magnetic resonance thermal imaging (MRTI) overcomes these limitations and has potential to produce less collateral tissue damage than current methods. Objective: To correlate multislice volumetric 2-D MRTI volumes with histologically confirmed regions of tissue damage in MRgRA NBTU. Methods: Seven swine underwent a total of 8 frontal MRgRA NBTU lesions. MRTI ablation volumes were compared to histologic tissue damage on brain sections stained with 2,3,5-triphenyltetrazolium chloride (TTC). Bland-Altman analyses and correlation trends were used to compare MRTI and TTC ablation volumes. Results: Data from the initial and third swine’s ablations were excluded due to sub-optimal tissue staining. For the remaining ablations (n = 6), the limits of agreement between the MRTI and histologic volumes ranged from −0.149 cm3 to 0.252 cm3 with a mean difference of 0.052 ± 0.042 cm3 (11.1%). There was a high correlation between the MRTI and histology volumes (r 2 = 0.831) with a strong linear relationship (r = 0.868). Conclusion: We used a volumetric MRTI technique to accurately track thermal changes during MRgRA NBTU in preparation for human trials. Improved volumetric coverage with MRTI enhanced our delivery of therapy and has far-reaching implications for focused ultrasound in the broader clinical setting.

Original languageEnglish (US)
Pages (from-to)907-915
Number of pages9
JournalInternational Journal of Hyperthermia
Volume38
Issue number1
DOIs
StatePublished - 2021
Externally publishedYes

Keywords

  • Brain metastases
  • focused ultrasound
  • magnetic resonance thermal imaging
  • magnetic resonance-guided robotically assisted delivery
  • needle-based therapeutic ultrasound

ASJC Scopus subject areas

  • Physiology
  • Physiology (medical)
  • Cancer Research

Fingerprint

Dive into the research topics of 'Predicting ablation zones with multislice volumetric 2-D magnetic resonance thermal imaging'. Together they form a unique fingerprint.

Cite this