Skip to main navigation Skip to search Skip to main content

Smart Offloading Boot System for Remote Patient Monitoring: Toward Adherence Reinforcement and Proper Physical Activity Prescription for Diabetic Foot Ulcer Patients

  • Catherine Park
  • , Ramkinker Mishra
  • , Davide Vigano
  • , Maurizio Macagno
  • , Stefano Rossotti
  • , Karen D’Huyvetter
  • , Jason Garcia
  • , David G. Armstrong
  • , Bijan Najafi

Research output: Contribution to journal › Article › peer-review

Abstract

Background: A critical factor in healing diabetic foot ulcers is patient adherence to offloading devices. We tested a smart offloading boot (SmartBoot) combined with a smartwatch app and cloud dashboard to remotely monitor patient adherence and activity. In addition, the impact of SmartBoot on balance, gait, and user experience was investigated. Methods: Fourteen volunteers (31.6±8.7 years; 64% female) performed natural activities (eg, sitting, standing, walking) with and without the SmartBoot for approximately 30 minutes. All participants completed balance tests, 10-meter walking tests at slow, normal, and fast pace while wearing the SmartBoot, and a user experience questionnaire. The accuracy of real-time adherence reporting was assessed by comparing the SmartBoot and staff observation. Center of mass (COM) sway and step counts were measured using a validated wearable system. Results: Average sensitivity, specificity, and accuracy for adherence and non-adherence were 90.6%, 88.0%, and 89.3%, respectively. The COM sway area was significantly smaller with the SmartBoot than without the SmartBoot regardless of test condition. Step count error was 4.4% for slow waking, 36.2% for normal walking, 16.0% for fast walking. Most participants agreed that the SmartBoot is easy to use, relatively comfortable, nonintrusive, and innovative. Conclusions: To our knowledge, this is the first smart offloading system that enables remote patient monitoring and real-time adherence and activity reporting. The SmartBoot enhanced balance performance, likely due to somatosensory feedback. Questionnaire results highlight SmartBoot’s technical and clinical potential. Future studies warrant clinical validation of real-time non-adherence alerting to improve wound healing outcomes in people with diabetic foot ulcers.

Original languageEnglish (US)
Pages (from-to)42-51
Number of pages10
JournalJournal of Diabetes Science and Technology
Volume17
Issue number1
DOIs
StatePublished - Jan 2023
Externally publishedYes

Keywords

  • adherence
  • diabetic foot ulcer
  • mobile health
  • remote monitoring
  • smart offloading boot

ASJC Scopus subject areas

  • Internal Medicine
  • Endocrinology, Diabetes and Metabolism
  • Bioengineering
  • Biomedical Engineering

Fingerprint

Dive into the research topics of 'Smart Offloading Boot System for Remote Patient Monitoring: Toward Adherence Reinforcement and Proper Physical Activity Prescription for Diabetic Foot Ulcer Patients'. Together they form a unique fingerprint.