Abstract
We have developed a prototype ultrahigh-resolution small-animal imaging system that utilizes a silicon detector with a multiple-pinhole aperture This system is suitable for imaging low-energy photons such as those from 1-125 decay (27.2-35.5 keV) or the X-rays from 1-123 decay (27.2-31.8 keV). The silicon detector is a double-sided strip detector of 300-micrometer thickness and 50-micrometer strip pitch, with 560 strips on one side and 1260 strips on the other. The readout utilizes VaTaGP2.2 ASICs from IDEAS ASA, Norway. When the detector is coupled with a multiple-pinhole aperture and operated in a synthetic-collimator configuration it should be possible to do tomographic image reconstructions without rotation of either the object or imager.
| Original language | English (US) |
|---|---|
| Article number | M4-1 |
| Pages (from-to) | 1984-1987 |
| Number of pages | 4 |
| Journal | IEEE Nuclear Science Symposium Conference Record |
| Volume | 3 |
| DOIs | |
| State | Published - 2003 |
| Event | 2003 IEEE Nuclear Science Symposium Conference Record - Nuclear Science Symposium, Medical Imaging Conference - Portland, OR, United States Duration: Oct 19 2003 → Oct 25 2003 |
ASJC Scopus subject areas
- Radiation
- Nuclear and High Energy Physics
- Radiology Nuclear Medicine and imaging
Fingerprint
Dive into the research topics of 'Ultrahigh-resolution small-animal imaging using a silicon detector'. Together they form a unique fingerprint.Cite this